Programmable UV-Vis spectrophotometer
The Camspec M508 is a single beam Programmable UV-Vis spectrophotometer it includes a 4 line x 20 character LCD display with a 9 button keypad. Stepper motors are used for control of wavelength and filters. It can store up to 200 methods and results for big analytical projects. Data output includes method number/name, wavelength, date/time, %T, Abs, concentration, unit selectable wavelength calibration and dark current measurement.
This Programmable UV-Vis spectrophotometer incorporates a 200 - 1000nm wavelength range and 5nm bandwidth with low stray light (0.1%T at 220nm and 340nm). Also it comes complete with a manual 4-cell changer with 2 x 10mm UV cells and 4 x 10mm glass cells. PC Application Software is an option for concentration curves, timescan and wavelength scanning.
View the PDF below to see a selection of the many accessories that may be used with the M508. Please enquire if your requirements are not listed.
Main Features of the Programmable UV-Vis spectrophotometer
- Stepper motors for wavelength and filters
- Stores up to 200 methods/results
- Concentration calibration: up to 8 standards and up to 3 replicates, linear or linear through zero
- Selectable wavelength calibration and dark current measurement
- 200 - 1000nm wavelength range
- 5nm bandwidth
- Low stray light (0.1%T at 220nm and 340nm)
- Large sample compartment (up to 100mm pathlength cells)
- 2 x 10mm UV cells and 4 x 10mm glass cells included
- Large selection of accessories including a manual 4 cell changer
- PC Application Software option for wavelength scanning, concentration curves and timescan.
See our Standards for UV-Vis spectrophotometers
Specification of the Camspec M508 Programmable UV-Vis spectrophotometer
Display | 4 lines x 20 character LCD LCD |
Light sources | Tungsten-Halogen and Deuterium |
Monochromator | 1200 lines/mm grating |
Detectors | Silicon Photodiode |
Wavelength Range | 200 - 1000nm |
Wavelength Accuracy | ± 2 nm |
Wavelength Repeatability | ± 1 nm |
Noise | < 0.001A @ 500nm 0A |
Zero Drift | < 0.003A per hour after warm-up |
Bandpass | 5 nm |
Stray Light | <0.1%T @ 220nm and 340 nm |
Photometric Accuracy | +/- 1% 0-2 A |
Photometric Range | -1 to 2.5A, 0-125%T and -9999C and +9999C |
Printer Interface | RS 232 for printer |
Computer Interface | USB for PC control |
Power Requirements | 110/120V, 220/230V, 50/60 Hz, 135VA |
Dimensions | 510 x 420 x 210 packed 645 x 535 x 370 |
Weight | 14Kg packed 17Kg |
User InterfaceThe user interface of the Camspec M508 programmable UV-Vis spectrophotometer includes Basic Mode (Abs, %T, Conc), Quantitative (concentration calibration curves), Wavelength Scanning, Kinetics (including timedrive), DNA/Protein, Multi-wavelength and Performance Validation. In Basic Mode, the result is continuously displayed (no need to “press to read”) PC ControlOur M508 has a USB output for the transfer of results to a PC, and another USB for PC control of the M508. Comprehensive Camspec Application Software is available, together with small PC programs, dedicated to specialist tasks. Click here to see all accessories to fit our range of spectrometers |
Click here to view the PDF file with the full specification
We are an ISO 9001:2015 registered company.