FEATURES

  • UNIQUELY HYBRIDIZED BWB dual CHANNEL, FOCUSED FOR HIGH ACCURACY POTASSIUM
    and sodium MEASUREMENTS

  • “IRS” (INTERNAL REFERENCE STANDARD) AVAILABLE

  • 4-20MA 2 WIRE OUTPUT COMPATIBLE WITH SCADA SOFTWARE

  • LIVE ONLINE MONITORING

  • CONSTANT CONDENSATION STREAM FEEDS WITH USE OF THE PROPRIETARY BWB COLLECTION CUP

  • BUILT-IN AIR COMPRESSOR; PIONEERED BY BWB TECHNOLOGIES IN 2007 AND PROVEN NOW WITH OVER 1000 UNITS IN GLOBAL FIELD USE

  • SOLUTIONS AND LABWARE INCLUDED; “JUST ADD GAS”, THE HALLMARK OF THE USER FRIENDLY BWB FLAME PHOTOMETER (FES) PROGRAM

  • SPECIALIZED CALIBRATION “READY TO USE” SOLUTIONS SUPPLIED AT 1, 5, 10 AND 20PPM

  • USER SELECTABLE DECIMAL PLACES

  • INTUITIVE USER INTERFACE FOR TRUE EASE OF USE

  • DISPLAY PROMPTS STEP BY STEP OPERATION

  • DATA SHARING VIA PC LINK WITH BWB’S
    FP/PC APP

Technical Data

Sample rate      3.5 – 7ml/min

Tubing materials      Silicone and Tygon®

Required desk space for Instrument       50(H) x 45 (W) x 45 (D) cm

For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.

Instrument size      Sugar Flame Photometer 51cm (H) x 38cm (W) x 41cm (D)

Shipping     62cm (H) x 55cm (W) x 47cm (D)

Weight      Sugar Flame Photometer – 15.3kg

Shipping      25kg

Optimal range     Single point calibration Na 0.1 – 60ppm  K 0.05 – 100ppm

  Multi Point Calibration Na 0.1 – 1000ppm  K 0.05 – 1000ppm

Reproducibility 

<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).

Chemical Interference  Na/K/Li = <0.5% to each other when equal in concentration at <100ppm

Limit of detection (LOD) and
Limit of quantification (LOQ) 

LOD                      LOQ

Na – 0.03ppm     Na – 0.1ppm

K – 0.02ppm       K – 0.05ppm

Time to stability  Less than 15 seconds after sample is introduced into the flame

Ions measured  

Potassium
Sodium

Interfaces  USB

4-20mA output capable of direct link to SCADA systems

Optional integrated printer

.CSV and .PDF generated reports and files via FP-PC software

Collection Cup providing 24/7 live automated sampling*

Recommended minimum warm up
time based on ambient temperature   21ºC – 40 minutes

Power requirements  100V – 250V AC at 50 or 60Hz automatically selecting

Fuel requirement    Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of O.4l/min

*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.

Readout  LCD, four line, alpha numeric, back lit.